A useful demonstration that biosensor readouts and wavelength-resolved PAM can be recorded together on attached leaves, but Fig. 3B is inconsistent with Fig. 3A at 25 µE, no optical or sensor-free controls exclude non-pH signal contributions, and the biological conclusions rest on n = 3 without statistics (Fig. 4: a single leaf).
Soundness
2/4
Presentation
3/4
Contribution
3/4
Confidence
3/5
✓ 11 confirmed✕ 4 discrepancy◐ 2 inconclusive○ 1 not checkable
Summary
The paper couples genetically encoded fluorescent biosensors with the ChloroSpec L1, a time- and wavelength-resolved chlorophyll-fluorescence spectrometer with programmable actinic light, to read subcellular physiology and PAM-type photosynthetic parameters simultaneously from attached Arabidopsis leaves. Stromal pHluorin2 is first compared between confocal imaging of leaf discs (monophasic alkalinization) and ChloroSpec on attached leaves (overshoot, partial re-acidification, second rise) (Fig. 1). The approach is then applied to cytosolic pH (pHluorin2), stromal MgATP2− (ATeam) and cytosolic NADH/NAD+ (Peredox-mCherry) (Fig. 2).
Two physiological experiments follow: a stepwise light series (25–800 µE, 5 min light / 5 min dark) in which the stromal pH response shows an onset overshoot at every intensity, a post-light undershoot from 200 µE upward and an amplitude that flattens at 600/800 µE (Fig. 3); and a ~4 h sinusoidal fluctuating-light regime in which stromal pH tracks irradiance in phase, its mean rises progressively alongside NPQ, and Y(II), NPQ, Y(NPQ) and Y(NO) differ in a time-dependent way between 685, 700 and 720 nm detection (Fig. 4, Fig. S3). The authors present the combination as a platform that links cellular bioenergetics to leaf physiology. One author is CTO and co-founder of the instrument manufacturer (declared).
Strengths
Practically valuable measurement combination
Figs. 1D–E, 4; Methods p.12–13
Recording a ratiometric biosensor and PAM parameters (Y(II), NPQ, derived yields) from the same attached leaf under programmable light is hard with confocal setups, which are limited to excised discs and short sessions. The 4-h fluctuating-light run (Fig. 4) shows the setup can follow a sensor signal and PAM parameters over hours; the protocol timing as described (15 min at 150 µE, 3.81 h sinusoid 50–248 µE, periods 1.5 → 10.6 → 3.2 min) is reproduced from the figure (check 13).
Several sensors and compartments shown on one platform
Results p.4–5; Fig. 2A–C
The same acquisition scheme works for stromal and cytosolic pHluorin2, stromal ATeam and cytosolic Peredox (Fig. 2). The described kinetics (cytosolic acidification below baseline at ~3 min with persistent post-light alkalinization; ATeam maximum ~1.5–2 min after onset and return at light-off; Peredox plateau within ~1.2 min, 44% of the response remaining after 10 min of darkness) match the digitised mean traces (check 12).
Stepwise and fluctuating-light descriptions match the figures
Figs. 3A–B, 4A–C
Digitisation reproduces the stated qualitative features: onset overshoot at every intensity and a post-light undershoot from 200 µE upward that grows with intensity (minima −0.019, −0.042, −0.063, −0.086 at 200–800 µE; check 03); pH in phase with irradiance in Fig. 4A (median within-cycle r = +0.91, 39/39 cycles positive, lag ≤ 1 pixel), rising cycle means (0.04 → 0.10), Y(II)685 falling 0.46 → 0.24 and NPQ685 rising 0.61 → 1.32 (check 04); Fig. 3B points, means and s.d. are internally consistent (check 02).
Wavelength-resolved PAM parameters are a distinctive dataset
Results p.7–8; Fig. 4B,C; Methods p.13
Reporting Y(II) and NPQ separately at 685, 700 and 720 nm under fluctuating light (Fig. 4B,C) is of interest to the chlorophyll-fluorescence and solar-induced-fluorescence communities. The stated orderings in both phases (first phase Y(II) 0.485/0.446/0.329, NPQ highest at 720 nm; late phase 700 > 685 > 720 for both) are reproduced from the figure (check 05), and the Y(NO)/Y(NPQ) statements attributed to the missing Fig. S3A largely follow from Fig. 4 via the Methods identities (checks 06, 07).
Candid discussion of platform differences, explicit formulas and declared COI
Discussion p.8–9; Methods p.13; p.14; References
The Discussion openly lists why confocal and ChloroSpec responses differ (light gradient through the leaf, uncross-calibrated light meters, wounding and gas exchange of discs) rather than forcing agreement, the PAM parameter definitions are given explicitly and are mutually consistent (Y(II)+Y(NPQ)+Y(NO) = 1, Y(NPQ)/Y(NO) = NPQ; check 07), the commercial interest of one author is declared, and the bibliography is clean (45/45 references exist; check 15).
Weaknesses
Fig. 3B amplitude at 25 µE is mathematically incompatible with the Fig. 3A mean trace
major · Fig. 3A,B; Results p.6 (check 01)
The amplitude is defined as the maximum after light onset minus the minimum after light-off (Results p.6; Fig. 3 legend). For any set of leaves, the mean of per-leaf amplitudes must be at least the amplitude of the mean trace. At 25 µE the Fig. 3A mean trace rises to +0.049 and falls to −0.010, a lower bound of 0.059 (0.042 even using only the late-light plateau), but Fig. 3B reports a mean of 0.028 (individual points ≈ 0.006, 0.023, 0.061). The bound holds at all other intensities. Either the two panels were computed from different data/windows or the amplitude was computed differently from the stated definition. Because the 25-µE point anchors the low end of the intensity dependence and the claim that the overshoot is intrinsic at low light, the authors should recompute Fig. 3B from the per-leaf traces, state the exact time windows, and provide the per-leaf time series.
No optical or sensor-free controls for the whole-leaf ChloroSpec signal
major · Figs. 1E, 2, 3A, 4A; Discussion p.10
ChloroSpec integrates fluorescence over the illuminated leaf area, and leaf optics change with light: chloroplast movement, zeaxanthin (ΔA505) and electrochromic (P515) absorbance changes near the 511-nm emission band, and NPQ-related absorbance changes. Ratioing at one emission wavelength cancels emission-side reabsorption only to first order; excitation-side changes at 405 vs 465 nm need not cancel. The overshoot is likened to P515 kinetics (Discussion p.10) without excluding that it partly is one. In Fig. 3A at 25 and 100 µE a thin upward spike also appears right after light-off (+0.038, +0.032), of similar size to the onset spike (check 03), which suggests a transition artefact. Recordings of wild-type (no sensor) leaves and of a pH-insensitive fluorescent protein in the stroma under identical protocols are needed before the overshoot and the Fig. 4 trends can be attributed to stromal pH.
Peredox read as tSapphire intensity only, but interpreted as NADH/NAD+
Methods (p.12) state that the Peredox signal is tSapphire intensity alone, without the mCherry normalisation that makes Peredox-mCherry ratiometric. An intensity signal is sensitive to expression level, focus/geometry, photobleaching and any light-induced change in leaf absorbance. The Fig. 2C response is a ~1.9-fold intensity rise (check 12) and is interpreted as reduction of the cytosolic NAD pool, as is the slow 'integration' under fluctuating light (Fig. S3B; abstract). Either acquire the mCherry channel, or show controls (sensor-free leaves, a non-responsive Peredox variant, or an intensity-matched constitutive FP) demonstrating that the intensity change is NADH/NAD+-specific; until then the NAD redox claims in the abstract should be softened.
Questions for the authors
How was the Fig. 3B amplitude computed (exact windows for the maximum after onset and the minimum after light-off, per leaf)? Can you provide the three per-leaf traces at 25 µE so that Fig. 3B can be reconciled with the Fig. 3A mean (mean-trace amplitude ≥ 0.059 vs reported mean 0.028)?
Do wild-type leaves without a sensor, or leaves expressing a pH-insensitive FP in the stroma, show any 405/465-nm ratio change (onset overshoot, light-off spike, slow drift) under the Fig. 3 and Fig. 4 protocols?
What do the other two leaves of the fluctuating-light experiment show for the progressive stromal alkalinization, the NPQ rise and the 685/700/720-nm orderings, and does a time-matched constant-light control with the same measuring schedule show a similar slow rise?
What is the sampling interval, the sensor-excitation photon flux at 405/465 nm, and the frequency of the saturating-flash sequence, and how much did the sensors photobleach over 4 h?
Has the stromal pHluorin2 ratio been calibrated in planta? Over what pH range is it close to linear, and could the flattening at 600/800 µE reflect sensor saturation?
Was the mCherry channel of Peredox-mCherry recorded at all? If so, does the tSapphire/mCherry ratio reproduce the Fig. 2C and Fig. S3B responses?
Verification
No data, code or supplementary figures were available, so 18 checks were run on what the PDF contains: the raster Figs. 1–4 were digitised by pixel colour (axis-calibrated, tolerances stated per check), the Methods formulas were tested symbolically and numerically, physical plausibility of the flash protocol and of the proposed PSI/reabsorption mechanism was modelled, and references were checked against Crossref and the cited articles. Most textual descriptions of Figs. 1–4 are reproduced (checks 02–07, 11–13), but Fig. 3B is inconsistent with Fig. 3A at 25 µE (check 01), the proposed mechanism for wavelength-dependent PAM parameters cannot produce the reported orderings (check 08), the '150 single-turnover flashes' are multi-turnover (check 09), and Mahon 2011 does not report the 387/482-nm maxima attributed to it (check 16). Normalisation and sensor-range questions remain inconclusive without raw data or calibration (checks 10, 14), and Figs. S1–S3 could not be examined (check 18). All Fig. 4 checks rest on a single representative leaf.
Mahon, M. J. (2011). pHluorin2: An enhanced, ratiometric, pH-sensitive green florescent protein. Adv. Biosci. Biotechnol. 2(3), 132–137
Cited (p.3) for pHluorin2 excitation maxima of 387 and 482 nm; the article reports peaks at 395 and 475 nm (emission 509 nm) and the values 387/482 do not occur in it (probably from the co-cited Cislaghi et al. 2026 preprint). The single-author paper is also cited as 'Mahon et al., 2011' on p.3 but 'Mahon, 2011' on p.2.
Suggestions
Recompute Fig. 3B from the per-leaf traces with explicitly stated time windows and reconcile it with Fig. 3A; deposit per-leaf time series (all figures) in a public repository.
Add sensor-free (wild-type) and pH-insensitive-FP controls recorded with the identical ChloroSpec protocols to separate pH signals from leaf-optical and measuring-light artefacts.
Record the mCherry channel of Peredox-mCherry (or add equivalent controls) before interpreting intensity changes as NADH/NAD+.
Show all three replicates of the fluctuating-light experiment with summary statistics, and add a time-matched constant-light and dark control with the same measuring and flash schedule to quantify drift and photobleaching.
Calibrate pHluorin2 in planta or relabel axes as normalised log ratios, and discuss the sensor's dynamic range near stromal pH 8.
Report the sampling interval, sensor-excitation intensities, saturating-flash frequency and duration, Fv/Fm (per detection wavelength), and describe the 130-µs pulse train as multi-turnover.
Randomise or reverse the intensity order in the stepwise series on separate leaves, or temper the 'dark-adapted' wording on p.10.
Add statistical tests or confidence intervals for the intensity dependence (Fig. 3B) and the fluctuating-light trends.
Include Figs. S1–S3 with the manuscript and fix the Mahon 2011 attribution and the author-contribution initials.
Ethics and integrity
Declared conflict of interest: A. R. Holzwarth is co-founder and CTO of ChloroSpec B.V., which makes the instrument evaluated here; the paper functions partly as an application note for that product, which makes independent controls and data availability more important. The AI-use statement (image generation for cartoons, language editing) is disclosed. No data-availability statement was found. No signs of fabrication were found; the Fig. 3A/3B inconsistency is reported as a discrepancy that an analysis or windowing error could explain. No instructions aimed at reviewers were found in the manuscript.
Scope of this review
Only the 21-page PDF was available: no raw or per-leaf data, no analysis code, and no supplementary figures (S1–S3), although the text relies on them for the CLSM high-light comparison, emission spectra and the fluctuating-light Peredox result. All quantitative checks on Figs. 1–4 come from digitising raster images (typical tolerance ±0.005 in log-ratio units, ±0.01–0.02 for PAM parameters), and Fig. 4 shows a single leaf. Sensor-range conclusions (check 14) use literature pKa and dynamic-range values because the paper gives no calibration. Wet-lab aspects (line construction, sensor localisation, instrument calibration) could not be assessed.
Fluctuating-light conclusions rest on one leaf and are confounded with elapsed time
major · Fig. 4A–C; Results p.6–8; Methods p.13
Fig. 4 is 'a representative trace from three independent repetitions' (Methods p.13); no replicate spread, summary statistics or per-leaf values are given for the progressive alkalinization, the NPQ rise or any wavelength ordering. The progressive rise of the pH ratio over ~4 h (cycle means 0.04 → 0.10; check 04) is also confounded with sensor photobleaching of the two pHluorin2 excitation states under repeated 405/465-nm excitation and 15,000 µE flashes, and with baseline drift, which in ChloroSpec dark recordings is 4–10× larger than in CLSM (0.05–0.13 vs 0.012; check 10). Show all three leaves, add a time-matched constant-light (e.g. 150 µE) and a dark control with the same measuring-light/flash schedule, and quantify the trends across replicates.
No statistics; trend claims exceed what n = 3 resolves
major · Fig. 3B; Fig. 2A; Results p.4–6; Methods p.13
No statistical test is reported anywhere. In Fig. 3B the overall trend is clear (Spearman ρ = 0.93), but none of the neighbouring steps above 100 µE is resolved (Welch p = 0.11 for 200→400, 0.30 for 400→600, 0.48 for 600→800 µE; check 02), so 'increased with light intensity across the range tested … increments became smaller at 600 and 800 µE' describes means, not demonstrated differences. Small features such as the 'brief alkalinization' in cytosolic pH (Fig. 2A) exceed the extrapolated pre-light dark trend by only ~0.02, half the dark-drift range (check 12). Report tests or confidence intervals and qualify the wording.
Ratios labelled and discussed as 'pH' without calibration; sensor saturation not considered
major · Results p.3, p.6; Discussion p.9–10; Figs. 1–4 axes
All axes read 'Stromal pH' or 'Cytosolic pH' but show log10 normalised 405/465 ratios; no in-situ calibration or pKa is given. With literature pKa ≈ 6.9–7.3, the log-ratio sensitivity at stromal pH 8.0 is only 0.13–0.39 of its maximum, and equal pH steps give increments shrinking by 25–35% per step (check 14). The flattening at 600/800 µE (Fig. 3B), attributed to limits on proton flux (Discussion p.9–10), could therefore arise from the sensor's titration curve alone. Calibrate in planta (e.g. nigericin/buffer clamps) or at least relabel axes as ratios and discuss the sensor's range.
Offered mechanism for wavelength-dependent PAM parameters does not explain the reported orderings
minor · Results p.7–8; Discussion p.11; Fig. 4B,C
The Results/Discussion attribute the wavelength differences to more non-variable PSI fluorescence above 700 nm and stronger reabsorption below 700 nm. A uniform reabsorption factor cancels in Y(II) and NPQ, and an additive PSI offset lowers both Y(II) and NPQ (check 08). Fitting the offset that lowers Y(II) from 0.485 (685 nm) to 0.329 (720 nm) predicts NPQ720 ≈ 0.29, but 0.549 is observed; the late-phase 700 > 685 ordering would require a negative PSI contribution at 700 nm. The dark Y(II) after light-off is also 0.10 higher at 700 than at 685 nm (check 05). The authors note the reversal but provide no mechanism; depth-dependent reabsorption with heterogeneous quenching, or state transitions, should be discussed or modelled, and Fv/Fm by wavelength reported.
Increasing-intensity order and 5-min dark gaps; 'dark-adapted' overstatement
minor · Results p.5–6; Discussion p.10; Fig. 3A
The six intensities were always given in increasing order with 5-min dark intervals after a 1-h dark adaptation, so intensity is confounded with cumulative light history and time. The Discussion states that 'every increase in irradiance was a step that began from a dark-adapted state' (p.10), yet the undershoot from 200 µE upward and end-of-dark levels of +0.004 to +0.017 (check 03) show the leaf had not returned to its initial state. Randomise or reverse the order on separate leaves, or rephrase.
The sampling interval, sensor-excitation LED intensities and the frequency/duration of the saturating-flash sequence are not stated, so the normalisation to 'five pre-light points' cannot be checked (check 10; Fig. 3A bars suggest ≈0.75 min per sample) and the ChloroSpec overshoot peak at ~0.4 min (Fig. 1E) is resolved by only one or two samples (check 11). The '150 single-turnover flashes' of 130 µs at 60,000 µE deliver ~7.8 µmol photons m⁻² each, i.e. ~3–14 hits per PSII, with 4-µs gaps far shorter than QA⁻ reoxidation, so the train is effectively a ~20-ms multi-turnover pulse (check 09). Fv/Fm is never reported although it is the standard quality check of the dark-adapted leaf.
Reversibility overstated; key comparisons rest on missing supplementary figures
After 10 min of darkness the stromal ratio retains 19% (CLSM) and 34% (ChloroSpec) of the light plateau (check 11); the text calls both responses reversible without noting the incomplete return for ChloroSpec. The argument that the platforms are 'complementary rather than equivalent' relies on CLSM data at 200/850 µE in Fig. S1, and the NAD redox fluctuating-light result on Fig. S3B; Figs. S1–S3 were not in the PDF (check 18). The new cytosolic pHluorin2 line is not characterised (localisation images, number of lines, expression).
Citation and attribution slips
minor · Results p.3; Introduction p.2; Author contributions p.14
The pHluorin2 excitation maxima of 387/482 nm are attributed to Mahon (2011), whose article reports 395/475 nm (check 16); the single-author Mahon paper is cited as 'Mahon, 2011' (p.2) and 'Mahon et al., 2011' (p.3). In the author contributions, Karin B. Busch appears as both 'KBB' and 'KB', and 'MH' is ambiguous between Minh Hoang and Michael Hippler ('MHip') (check 17).
Traces are log10 of ratios normalised to the mean of the five pre-illumination points (so the plotted mean trace averages ≈0 over those points); pre-light baselines are stable enough to interpret the light responses
For every panel some window length gives a pre-light mean within ±0.005 of 0 (1C, 1E: any W; 2A: only W ≥ 6.1 min; 2B: 0.4–3.5 min; 2C: 0.6–5.0 min; 3A: 0.5–5.1 min), so the stated normalisation cannot be refuted without the sampling interval. The Fig 3A s.d.-bar spacing suggests ≈0.75 min per sample (IQR 0.60–0.90; five points ≈3.8 min), for which Fig 2A would average about +0.011 rather than ≤0 (weak hint only). Pre-light drift range of the mean trace (darkness, before any light): 1C 0.012 (CLSM) vs ChloroSpec 1E 0.058, 2A 0.048, 2B 0.128, 2C 0.080, 3A 0.049; light-response sizes 0.110, 0.147, 0.084, 0.305, 0.309 and 0.042 (3A, 25 µE). The ChloroSpec dark drift is 4–10× larger than with CLSM and is about as large as the 2A (cytosolic pH) and 25-µE responses, so their small features (e.g. the 'brief alkalinization' in Fig 2A) are not clearly separated from baseline drift.
Digitised the mean traces; averaged them over the last W min before light-on for W = 0.3 min … full dark period (sampling interval not reported); measured pre-light drift range and light-response size; estimated the sampling interval from s.d.-bar spacing in Fig 3A
Fig 1: CLSM (leaf discs) shows a monophasic rise to a plateau and return close to baseline; ChloroSpec (attached leaves) shows an overshoot, partial re-acidification and second rise; both responses reversible
1C: no overshoot (peak − dip = −0.001), plateau +0.109, 90% of plateau after 3.0 min. 1E: early peak +0.146 at 0.36 min after light-on, dip +0.063 at 2.15 min (overshoot 0.083), plateau +0.147. Shapes match the description. Reversibility is partial on both platforms after 10 min of darkness: residual +0.021 (19% of plateau) for CLSM, described as 'close to' baseline, and +0.051 (34%) for ChloroSpec, which the text calls reversible without noting the incomplete return. The ChloroSpec peak is reached within ~0.4 min, i.e. within one or two samples at the ≈0.75-min interval estimated in check 10, so the overshoot's time course is barely resolved.
Digitised Fig 1C and 1E mean traces; early peak (first 1.5 min of light) vs dip (1.5–4 min), plateau (last 2 min of light), post-light residual (last 3 min) as % of plateau
Fig 2 kinetics: cytosolic pH brief alkalinization, acidification below baseline after ~3 min, persistent alkalinization after light-off; stromal MgATP2− max after ~2 min and return to about pre-illumination at light-off; cytosolic NADH/NAD+ rises within ~3 min and does not return in darkness
2A: trace goes below 0 at +1.7 min, minimum −0.042 at +2.8 min, end-of-light +0.101, post-light +0.103 (persists) ✓. But the 'brief alkalinization' spike (+0.049 at +0.17 min) exceeds the extrapolated pre-light dark trend (+0.009 per min, already +0.022 before light-on) by only +0.020, about half the dark drift range (0.048). 2B: light maximum +0.287 at +1.5 min (text: ~2 min), late light +0.279, 1–3 min after light-off +0.002 vs pre-light −0.016 ✓; the 5-min pre-light trace ranges −0.033 to +0.095. 2C: 90% of plateau at +1.2 min, plateau +0.281 = 1.93-fold rise in tSapphire intensity, 44% of the response left after 10 min of darkness ✓; there is also a step of −0.019 within ~1 min of light-off. Because Peredox is read from tSapphire intensity only (Methods p.12), a ~2-fold intensity change is also open to optical explanations (leaf optics, chloroplast movement, excitation-side absorbance), which the paper does not control for.
Digitised the three mean traces; extracted peak times, minima, plateaus and post-light levels; compared the 2A 'brief alkalinization' with the extrapolated pre-light dark trend
Fluctuating-light protocol: 15 min at 150 µE, ~3.8 h sinusoid 50–250 µE with period ~2 → ~11 → ~3 min, ~15 min final dark, ~4 h in total; first constant phase at 10–25 min; NADH/NAD+ maximum at 45–50 min = 20–25 min after modulation start
Constant phase at 150 µE until 25.6 min (≈15.6 min from the 10-min light-on); modulation 25.6–254.1 min = 3.81 h; irradiance range 50 (axis floor) – 248 µE, time-averaged 157 µE; periods 1.5/2.2/1.5 min at the start, maximum 10.6 min at ≈130 min, 2.9/3.2/3.2 min at the end; total illumination 4.07 h; final dark shown for 13.0 min (text: about 15 min). 45–50 min is 19–24 min after modulation start (text: 20–25 min). All within digitisation tolerance (±0.3 min). Fig S3B itself (Peredox) is not available, so the NADH/NAD+ maximum time could not be checked against data.
Digitised the irradiance envelope of Fig 4 (white/grey boundary, right axes); onset, constant level, modulation start/end, extremes, period sequence from irradiance minima
The pHluorin2 405/465 ratio reports stromal pH across the light-induced range, and the smaller amplitude increments at 600 and 800 µE reflect physiology (limits on proton flux)
At pH 8.0 (typical illuminated stroma) the log-ratio sensitivity is only 0.13–0.39 of its maximum, and 0.07–0.24 at pH 8.25. Equal +0.2 pH steps from pH 7.4 give log-ratio increments that shrink by ~25–35% per step (e.g. pKa 7.1, DR 10: 0.049, 0.035, 0.024, 0.016). The flattening of Fig 3B at 600/800 µE can therefore arise from the sensor's titration curve alone; with no calibration in planta the data cannot separate this from physiological saturation. The y axes labelled 'Stromal pH' show log10 normalised ratios, not pH.
Two-state ratiometric titration model with literature pKa 6.9–7.3 (no pKa or in-situ calibration given in the paper; reviews list 7.1 for pHluorin2) and dynamic range Rmax/Rmin 3–10; relative sensitivity d log10R/dpH over pH 7.0–8.5 and log-ratio increments for equal +0.2 pH steps
All 45 references exist with the stated metadata, including the recent ones (Cislaghi 2026, Cainzos 2026, Hipsch 2025, Zheng 2025)
38/45 match on title, year and first author. 7 flagged, all benign: Franck 2005, Kaiser 2018, Kramer 2011 and Terashima 2011 differ by one year (Crossref online date vs print year); Müller 2001 is a Unicode-normalisation artefact; Klughammer & Schreiber 2008 and Schreiber & Klughammer 2008 are Walz PAM Application Notes, not indexed by Crossref (grey literature, exist). Cislaghi 2026 and Zheng 2025 resolve to bioRxiv preprints by overlapping author groups, which carry several methodological claims (pHluorin2 spectra, CLSM biphasic responses at 200/850 µE, brief NAD oxidation).
tools/check_refs.py (Crossref by DOI or bibliographic query) re-run in this step; script compares found title/year/first author with the reference text
Load-bearing citations support the statements attributed to them: Mahon 2011 (pHluorin2 excitation maxima 387 and 482 nm), Stroch 2005 (660–710 nm detection gives higher NPQ and PSII efficiencies than far-red detection), Voon 2018 (negligible ATP import into mature chloroplasts), Scherschel 2024 (Peredox shows reduction of cytosolic NAD pool in illuminated leaf discs), Steinbeck 2020 (dark–light NAD redox dynamics)
Mahon 2011: abstract and full text report a bimodal excitation spectrum 'with peaks at 395 and 475 nm and an emission maximum at 509 nm' (for ratiometric pHluorin); the strings 387 and 482 do not occur in the article text, so the 387/482 nm maxima attributed to Mahon (p.3, cited as 'Mahon et al., 2011' though single-author) are not from that source (probably from the co-cited Cislaghi et al. 2026 preprint). Minor. The others support the statements: Stroch 2005 – detection 'in the range of 660–710 nm led to pronounced increase of NPQ, SV0, and the PS2 photochemical efficiencies (FV/FM and FV′/FM′)'; Voon 2018 – 'the import of ATP into mature chloroplasts is impeded …', transport 'undetectable by our methods'; Scherschel 2024 – illumination of mature leaf discs 'revealed a pronounced reduction of the cytosolic NAD pool as monitored by Peredox'; Steinbeck 2020 – dark-to-light transitions revealed plasticity of the cytosolic NAD redox status and redox coupling between compartments.
Europe PMC REST (abstracts by DOI, open-access full text), NCBI PMC full text (Mahon), Photosynthetica open-access page (Stroch); printed matching sentences
Every in-text citation appears in the reference list and vice versa; author-contribution initials are consistent
45 distinct author–year citations, all in the reference list; no uncited references. Minor presentation issues: the single-author Mahon (2011) is cited both as 'Mahon, 2011' (p.2) and 'Mahon et al., 2011' (p.3); in the contributions, Karin B. Busch appears as 'KBB' (funding, administration, editing) and 'KB' (supervision), and 'MH' (methodology) could be either Minh Hoang or Michael Hippler (who is 'MHip' elsewhere).
Regex extraction of author–year citations from the body text, matched against the 45 reference keys; mapping of contribution initials to the author list
Statements resting only on supplementary figures: CLSM shows a biphasic stromal pH response only at 200 and 850 µE (Fig S1); full emission spectra recorded during the intensity series (Fig S2A–C); cytosolic NADH/NAD+ under fluctuating light peaks at 45–50 min and does not follow individual fluctuations (Fig S3B)
The text cites Figs S1, S2A–C, S3A and S3B, but the 21-page PDF has no supplementary captions or pages (4 embedded images = Figs 1–4) and no supplement was supplied. The CLSM-vs-ChloroSpec comparison at higher intensities (used to argue that the platforms are 'complementary rather than equivalent') and the Peredox fluctuating-light result cannot be examined. Fig S3A quantities were partly reconstructed from Fig 4 (check 06). Needed: the supplementary figures and, ideally, the underlying per-leaf time series.
Listed all in-text references to Fig. S*; searched the PDF text and embedded images for supplementary pages; checked the files supplied with the request
Fig 3B amplitudes (max after light onset minus min after light-off; mean ± s.d., n = 3) are derived from the same data as the Fig 3A mean traces
Fig 3B is internally consistent (digitised points 25 µE: 0.006, 0.023, 0.061; mean 0.030 vs dashed line 0.028; s.d. 0.028 vs half error bar 0.022; other intensities agree within 0.006). But at 25 µE the Fig 3A mean trace rises to +0.049 after onset and falls to −0.010 after light-off, a lower bound of 0.059 (0.080 including blended spike pixels; 0.042 even using only the late-light plateau +0.031 and dark minimum −0.011), whereas Fig 3B gives a mean amplitude of 0.028, i.e. 2.1× (1.5× for the spike-free bound) smaller than mathematically possible if both panels come from the same three leaves with the stated definition. At 100–800 µE the inequality holds (Fig 3B means 0.102, 0.171, 0.227, 0.251, 0.266 vs trace bounds 0.053–0.150). Tolerance 0.005 units.
Digitised the Fig 3A mean line (pure colour 166,3,219; axes calibrated from tick marks, 103.25 px per 0.1, 6.633 px/min) and the Fig 3B points, dashed means and error bars. Because mean_i(max_i − min_i) ≥ max_t(mean) − min_s(mean) over the same windows, the Fig 3B mean must be ≥ the amplitude of the Fig 3A mean trace. Conservative bounds: max restricted to the light window, min to the following dark window, half line width removed; plus a spike-free bound (late-light plateau minus lowest line centre in the following dark).
Amplitude of the stromal pH response increased with light intensity across 25–800 µE, with smaller increments at 600 and 800 µE; plotted means/s.d. match the individual points
Digitised means 0.030, 0.102, 0.171, 0.227, 0.252, 0.266 match the dashed lines (0.028, 0.102, 0.171, 0.227, 0.251, 0.266; tolerance 0.005). Means increase monotonically; increments 0.072, 0.070, 0.055, 0.025, 0.014, so the stated flattening is visible. Overall trend is strong (Spearman ρ = 0.93, p = 2e-8), but with n = 3 neighbouring steps above 100 µE are not resolved (Welch p = 0.11 for 200→400, 0.30 for 400→600, 0.48 for 600→800; 400 vs 800 p = 0.17). The paper reports no statistical test, so 'increased … across the range tested' is a description of means, not a demonstrated increase at the top of the range. (Internal consistency with Fig 3A at 25 µE fails; see check 01.)
Digitised the 18 Fig 3B points, dashed means and error bars; recomputed means, s.d., increments; Welch t-tests between neighbouring intensities, Spearman trend and one-way ANOVA
Overshoot at light onset at every intensity incl. 25 µE; at light-off pH fell below the pre-illumination level from 200 µE upwards (undershoot, increasing with intensity) and relaxed towards baseline within 5 min
Mean-trace overshoot (strict) +0.018, +0.014, +0.012, +0.007, +0.012, +0.020 at 25–800 µE (loose, incl. thin spikes: +0.038 to +0.059); post-light minima −0.010, −0.001, −0.019, −0.042, −0.063, −0.086, so the undershoot from 200 µE and its growth with intensity are reproduced. At 25 µE the trace also dips to −0.010, which is within the pre-light dark drift (trace at end of first dark ≈ −0.01). End-of-dark levels +0.004 to +0.017 (relaxation 'towards' baseline holds). Caveats: the overshoot is small relative to the dark drift in the first dark period (+0.06 → −0.01), and at 25 and 100 µE the mean line also shows a thin upward spike right after light-off (+0.038 and +0.032 above the end-of-light level, loose tolerance) of similar size to the onset spike, which suggests that part of the onset 'overshoot' may be a transition artefact; an optical control would be needed to rule this out. Thresholds: overshoot > 0.005, undershoot < −0.005.
Digitised Fig 3A mean line; per light phase: peak in first 1 min minus plateau (last 1 min of light); lowest point in following dark vs 0 and vs preceding dark level; level at end of following dark; strict (pure colour) and loose (blended thin-spike) tolerance
Fluctuating light: stromal pH in phase with irradiance; mean pH of oscillations rises progressively; no overshoot after irradiance increases; pH drops below initial value after light-off; Y(II) inverse to irradiance and declining, NPQ in phase and rising
pH vs irradiance: median within-cycle r = +0.91 (39/39 positive), best lag −0.25 min (one pixel; in phase). Cycle-mean pH rises from 0.052/0.042/0.036 (first cycles) to 0.098/0.095/0.094 (last), slope +0.020 per h; by thirds 0.038, 0.083, 0.100, so most of the rise occurs in the first two-thirds. No overshoot on rising half-cycles (median max-during-rise minus value at irradiance peak +0.001; 0/38 cycles > 0.02). After light-off pH falls to −0.028 (below the 0 pre-illumination reference) and recovers to +0.009 by 267 min. Y(II)685: r = −0.98, cycle means 0.46 → 0.24 (−0.047 per h); NPQ685: r = +0.97, 0.61 → 1.32. Caveat: Fig 4 is a single leaf ('representative of three'); no time-matched constant-light or dark control separates light history from elapsed time/sensor drift.
Digitised irradiance envelope (white/grey boundary, max over 3 panels), purple pH trace (Fig 4A) and 685-nm Y(II)/NPQ traces (black, Fig 4B,C); 39 cycles delimited by irradiance minima; within-cycle correlation and best-lag; linear trend of cycle means; rise-phase overshoot metric
Detection-wavelength orderings: 10–25 min Y(II) 685 > 700 > 720 nm and NPQ highest at 720 nm; later part Y(II) and NPQ highest at 700 nm and lowest at 720 nm
First phase: Y(II) 0.485 / 0.446 / 0.329 (685/700/720), NPQ 0.421 / 0.377 / 0.549 (720 highest). Late modulation (150–254 min): Y(II) cycle maxima 0.339 / 0.400 / 0.280, minima 0.232 / 0.263 / 0.151; NPQ maxima 1.423 / 1.584 / 1.145, minima 0.970 / 1.109 / 0.830, so 700 > 685 > 720 for both. NPQ720 maxima drop below NPQ700 from ≈83–107 min. Also: after light-off (257–267 min) Y(II) is 0.449 / 0.549 / 0.388, i.e. 700 nm exceeds 685 nm by 0.10, a large wavelength effect on a dark-state yield. All from one leaf; no replicate spread or statistics are shown for any ordering. Digitisation tolerance ≈ ±0.01 (Y(II)) and ±0.02 (NPQ); overlapping traces can hide early-modulation orange peaks (not used for the stated claims).
Digitised black (685), orange (700), green (720) traces of Fig 4B,C; medians over 12–24 min; per-cycle maxima and minima summarised for 26–100 and 150–254 min; final dark 257–267 min
Fig S3A (not provided): Y(NPQ) highest at 720 nm in the first phase; later Y(NPQ) and Y(NO) highest at 720 nm and lowest at 700 nm; Y(NPQ) increases while Y(NO) stays nearly constant
First phase (12–24 min): Y(NPQ) 0.156 / 0.153 / 0.240 (685/700/720) → highest at 720 ✓. Late (150–254 min), matched states: Y(NO) 0.313 / 0.285 / 0.401 (high light) and 0.336 / 0.296 / 0.395 (low light) → highest at 720, lowest at 700 ✓. Y(NPQ) 0.451 / 0.452 / 0.460 (high) and 0.325 / 0.311 / 0.326 (low): 720 is marginally highest and 700 lowest only at low light, but the differences (≤ 0.015) are within the propagated digitisation error (≈ ±0.015), so this part of the statement is not resolvable from Fig 4 and rests on Fig S3A, which was not available. Time course at 685 nm: Y(NPQ) 0.35 → 0.45 (high light) and 0.17 → 0.33 (low light) from 26–100 to 150–254 min, while Y(NO) 0.325 → 0.313 and 0.345 → 0.336 (nearly constant) ✓. Single leaf, no replicate statistics.
Derived Y(NO) = (1 − Y(II))/(1 + NPQ) and Y(NPQ) = 1 − Y(II) − Y(NO) (identities from the Methods definitions, check 07) from digitised Fig 4B,C: per-column within windows, and from matched per-cycle envelopes (irradiance maxima: Y(II) min with NPQ max; minima: the reverse)
Y(II)+Y(NPQ)+Y(NO) = 1 and Y(NPQ)/Y(NO) = NPQ exactly (symbolic); hence Y(NO) = (1−Y(II))/(1+NPQ), used in check 06. Max numeric deviation 8.9e-16. These are the simplified Kramer/Klughammer–Schreiber forms; they reduce to Fv/Fm and NPQ = 0 at the dark-adapted reference. Fv/Fm itself is never reported, although it is the standard quality check for the dark-adapted leaf and is available from the stated protocol.
sympy simplification of the yield sum and of Y(NPQ)/Y(NO) − NPQ; 10,000 random numeric instances with Fs < Fm′ ≤ Fm
Wavelength differences in PAM parameters are expected because emission > 700 nm contains more non-variable PSI fluorescence and emission < 700 nm is more strongly reabsorbed
First phase: the PSI offset needed to lower Y(II) from 0.485 (685 nm) to 0.329 (720 nm) is c = 0.33·Fm, which predicts NPQ720 = 0.29, but 0.549 is observed (NPQ is highest at 720 nm, opposite to the offset prediction). Late phase: to turn 685-nm into 700-nm values requires negative offsets (c = −0.049/−0.042 at high light, −0.077/−0.064 at low light), i.e. less PSI at 700 than at 685 nm. Only '720 nm lowest' in the late phase fits the stated mechanisms. The explanation is therefore incomplete as given; the authors note the reversal but offer no mechanism (e.g. depth-dependent reabsorption combined with heterogeneous quenching across leaf layers, or time-varying PSI contribution through state transitions). This is a model-level test of an interpretation, not of the data.
Minimal model F_L = a_L·F_PSII + c_L: sympy shows a multiplicative factor (uniform reabsorption) cancels in Y(II) and NPQ, and an additive PSI offset c ≥ 0 lowers both Y(II) and NPQ (dY/dc = (Fs−Fm′)/(Fm′+c)², dNPQ/dc = (Fm′−Fm)/(Fm′+c)²). Offsets fitted to the digitised Fig 4 values (check 05)
Each measuring cycle applied a burst of 150 'single-turnover flashes' (130 µs, 60,000 µE, 4 µs dark intervals) followed by a 15,000 µE multi-turnover flash
Each pulse delivers 7.8 µmol photons m⁻² (1,170 µmol m⁻² per burst; burst duration 20.1 ms, duty cycle 0.97). Expected hits per PSII per pulse: median 4.5 (95% range 2.8–7.3) by route 1; 4.7–14.1 by route 2 (P(≥2 hits) ≥ 0.95). The 4 µs gaps are 0.2–2% of QA⁻ reoxidation times (0.2–2 ms). The pulses are therefore not single-turnover in the usual sense; the 150-pulse train acts as a ~20-ms multi-turnover pulse. This is a terminology/description issue (assumption-dependent, but the conclusion holds over the whole range); it also matters for how much extra light the sensors receive per measuring cycle, which cannot be evaluated because the cycle interval and the MT-flash duration are not reported.
Photon dose per pulse; hits per PSII per pulse from (1) leaf Chl content, absorptance, PSII share and Chl/PSII (Monte Carlo over literature ranges) and (2) functional cross-section 1–3 nm²; comparison of 4 µs gaps with QA⁻ reoxidation times