A significant step for cryo-ptychographic SPA: the reported FSC resolutions (3.3–4.1 Å) reproduce from the plotted curves and the rCTF physics checks out in simulation. However, the stated polar-decomposition step cannot recover the defocus sign used in most datasets, and there is no code or raw data. The fractionation gain and model-building claims also lack controlled comparisons and standard validation.
Soundness
3/4
Presentation
2/4
Contribution
3/4
Confidence
4/5
✓ 9 confirmed✕ 8 discrepancy○ 3 not checkable
Summary
The paper describes an end-to-end workflow for cryo-electron ptychography with single-particle analysis (SPA). Before reconstruction, it estimates the probe defocus and twofold astigmatism, including a smooth local defocus landscape, by fitting a 'Ronchigram CTF' (rCTF) to power spectra of the out-of-focus shadow images. It estimates the scan pitch, shear and rotation from averaged cross-correlations of neighbouring Ronchigrams. With these parameters as initialisation, iterative single-slice ptychography (TorchSlice) gives phase images that go straight into cryoSPARC without CTF correction. An optional reference-based refinement then estimates per-particle defocus.
Applied to TMV, horse spleen apoferritin and HBV capsids (Krios G4, 300 kV, 4 mrad, Arina detector), the workflow reaches 3.7, 4.1 and 4.5/4.7 Å. Reprocessing the public EMPIAR-12236 mouse apoferritin dataset improves its result from the published 5.8 Å to 3.7 Å. The central new element is dose fractionation in 4D-STEM: three low-dose scans of the same TMV area are reconstructed separately, then motion-corrected and dose-weighted. This gives a 3.3 Å map (5,330 particles), into which ModelAngelo builds an atomic model (0.71 Å RMSD to PDB 6SAE). The per-fraction maps (3.4/3.9/4.4 Å), layer-line fading and empirical dose weights document radiation damage across the fractions.
Strengths
Physically sound self-calibration of probe and scan from the data itself
Sec. 2.2, Figs 2–3; Methods 6.4–6.5, Eqs (1)–(8)
The workflow estimates defocus/astigmatism (Eq. 2) and scan geometry (Eq. 8) from the shadow image at low dose, with no calibration specimen, by averaging thousands of Ronchigram power spectra and cross-correlations. This removes a real practical barrier. Our wave-optical simulations confirm the two physical premises. The rCTF zeros fall at τn=nλC1 to within 0.27%, and the elliptical form with the full matrix holds under astigmatism, to within 0.02 in ring index (check 05). The cross-correlation peak lag equals to within 0.4%, and Fourier transforming the peak-centred cross-correlation recovers the same power spectrum (check 06). Reading the power spectrum off the cross-correlation peak (Fig. 3c–d) is a neat way to suppress background.
Weaknesses
Polar decomposition as described cannot recover the defocus sign used in most datasets
major · Methods 6.5, Eqs (8), (10), (11), L1094–1106
Methods 6.5 separates GR−1=QH−1 (Eq. 11) into a rotation and the inverse defocus matrix by polar decomposition. Polar decomposition always returns a positive semidefinite factor. Our numerical test (check 07, 200 random rotations per case) recovers and exactly only for . For nm it returns nm and a rotated by 180° (). For it returns and an improper (), which is not a rotation. Most datasets were recorded at negative defocus (TMV −190…−800 nm, apoferritin −220…−920 nm, DF-TMV −260…−530 nm). The rCTF power spectrum is also blind to the sign (check 05). Some extra constraint, such as a known camera orientation, must therefore have been used, and the paper does not state it. Separately, Eq. (10) contradicts Eq. (11) whenever (relative error 0.30 for a 0.3 rad rotation), because Eq. (8) omits the scan–camera rotation. Since automated parameter estimation is a headline contribution, the method must be stated so that it works as written. Fix: give the sign/orientation convention explicitly, correct Eqs (8)/(10), and say how the strongly astigmatic case is handled.
Questions for the authors
How is the defocus sign (and the 180° ambiguity of the rotation) fixed in practice when separating GR−1=QH−1, given that polar decomposition returns ∣H and fails for ? Is the camera–scan rotation calibrated separately?
Verification
We ran 17 scripted checks and recorded 3 more as not checkable: 9 confirmed, 8 discrepancies, 3 not checkable. Vector FSC curves, defocus maps, loss curves, line scans and dose-weight curves were extracted directly from the PDF. The headline resolutions reproduce to within 0.1 Å (3.69, 4.14, 3.63, 3.31, 3.34 Å). The rCTF and cross-correlation physics (Eqs 2, 8) were confirmed by independent wave-optical simulation to within 0.4%. The equal-N dose-fractionation gain (~0.41 Å) is consistent with the claim once the Methods comparator is used. The discrepancies are: the polar-decomposition step fails for negative or strongly astigmatic defocus; the HBV resolution is reported as both 4.7 and 4.5 Å; the total dose is given as 25/24/23 e⁻/Ų; the per-particle height claims are overstated; the Fig. 6f layer-line positions and Fig. 6g cut-offs are off; there is undocumented particle bookkeeping; and an exposure pairing from [53] is reversed. Scan-geometry values, the model RMSD and the iDPC/CTEM comparisons could not be checked because no code, raw data or released maps are available.
Resolves (doi 10.1007/978-3-030-57077-4_10), but it is a textbook chapter in a TensorFlow book, not the primary PyTorch reference (Paszke et al., NeurIPS 2019). The PDF also prints the publisher field as '???'. The wrong work is cited for the software used.
[53] Fromm et al., Seeing tobacco mosaic virus through direct electron detectors, J Struct Biol 189, 87 (2015)
Cited at L561-565 for CTEM motion of '4.2 Å and 2.9 Å' with acquisition times '(1 s and 4.4 s respectively)'. In [53] the 4.2 Å motion belongs to the K2 Summit dataset with a 4.4 s exposure and the 2.9 Å motion to the Falcon II dataset with a 1 s exposure, so the pairing is reversed (check 15). The 3.4 Å resolution (3.35 Å in [53]) and the 20 e−/Ų dose are correctly cited.
[25] Küçükoğlu et al., Low-dose cryo-electron ptychography of proteins at sub-nanometer resolution, Nat Commun 15, 8062 (2024)
Exists and supports the cited numbers (apoferritin 5.8 Å from 11,552 particles, Phi92 sheath 8.4 Å, a TMV reconstruction at 6.4 Å). Minor: L43 gives the range as '5.8 and 8.4 Å', omitting the 12.1 Å Phi92 capsid map in the same work.
[39] Dushimineza et al., Imaging ionic displacements across inclined ferroelectric domains by a reciprocity approach to 4D-STEM, Small (in print)
Listed as 'in print', but the DOI 10.1002/smll.75386 resolves on Crossref to the published article (Small, 2 Sept 2026). Volume and pages should be updated.
The author list is garbled. Crossref metadata stores the first author as given 'Hoelen L.' / family 'Lalandec Robert', and the bibliography renders this as two authors, 'Lalandec Robert, Hoelen L.', with inconsistent name formatting.
[54] Weis, Beckers, von der Hocht, Sachse: Elucidation of the viral disassembly switch of tobacco mosaic virus, EMBO Rep 20, e48451 (2019)
Suggestions
State the sign/orientation convention used to resolve GR−1=QH−1, correct Eqs (8) and (10) to include the scan–camera rotation, and describe how the ∣ case is handled.
Ethics and integrity
No integrity concerns were found. The figure data we extracted are internally consistent with the reported values, apart from the slips noted above. We found no embedded instructions aimed at reviewers. The manuscript has no competing-interests statement. Several authors develop the TorchSlice software used here, which should be declared.
Scope of this review
Only the main PDF was available. The Supplement (Suppl. Figs 1–11, Suppl. Note 1, Suppl. Table 1) was not provided. There is no code (TorchSlice and the preprocessing are unreleased) and no raw 4D-STEM data for the new datasets, and all EMD/PDB entries are on hold. We could therefore not rerun reconstructions, inspect maps or models, verify the scan-geometry estimates for the EPFL dataset, or check the iDPC and CTEM comparisons. Figure values were obtained by extracting vector graphics from the PDF. The method's physics was checked with independent simulations, not with the authors' implementation.
Large, verifiable resolution improvement for ptychographic SPA
Sec. 2.6.1, 2.7; Fig. 4e, 4i; Fig. 6e
The FSC=0.143 crossings extracted from the vector FSC curves match the stated resolutions to within 0.1 Å. TMV: 3.69 Å (claimed 3.7). Horse apoferritin: 4.14 Å (4.1). Mouse apoferritin: 3.63 Å (3.7). Dose-fractionated TMV: 3.31 Å (3.3) (checks 01, 02). On EMPIAR-12236, a ResLog estimate shows that doubling the particle count explains only 21–33% of the 5.8 → 3.7 Å gain, and the voxel correction explains 5.8 → 5.67 Å. Most of the improvement therefore comes from the new processing, as claimed (checks 03, 14). The previous ptychographic SPA baselines (5.8 Å, 8.4 Å, 6.4 Å in [25]) are cited correctly (check 15).
First demonstration of dose-fractionated 4D-STEM ptychography with a coherent damage analysis
Sec. 2.4, 2.7; Figs 4g–i, 6; Methods 6.11.1, 6.11.5
Three ~8 e⁻/Ų scans are each reconstructed, then motion-corrected and dose-weighted in a standard cryo-EM pipeline. Three independent readouts agree on the damage ordering: per-fraction FSC (3.44/3.85/4.40 Å extracted vs 3.4/3.9/4.4 Å claimed, check 02), layer-line fading (Fig. 6f), and empirical dose weights. The dose weights decrease monotonically, with DC weights of 0.363/0.336/0.300 (check 11). Methods 6.11.1 contains a full 5,683-particle single-scan map at 3.7 Å. Measured against it, the equal-particle-number gain is ≈0.41 Å, consistent with the stated 'approximately 0.4 Å' (check 03).
Honest treatment of per-particle defocus refinement
Sec. 2.5, Fig. 5c; Sec. 2.6.1 L448–454; Methods 6.10
Fig. 5c shows openly that the ptychographic reconstruction loss is a poor focus criterion. Its global minimum lies at the +100 nm edge of the scanned range (check 09). The authors replace it with a reference-based MSE that uses the matching gold-standard half-map (Methods 6.10, L1297–1300), which avoids half-set leakage. They also report candidly that per-particle refinement gives no resolution gain for TMV, and explain this with the depth of focus. The stated 221 nm corresponds to the 1.8λ/α2 convention (221.5 nm, check 12).
Quantitatively consistent local-defocus and acquisition figures
Fig. 2d–e, L209–214; Methods 6.2, 6.8
The figures we could check hold up numerically. The Fig. 2d ring-detection limits are 24.4/29.7 Å (claimed 24/30 Å). The Fig. 2e defocus landscape spans 0.722–0.914 µm, which bounds the planar tilt to 14.9–20.6° and contains the reported 17.4° (check 08). Scan time is 1024²/100 kHz = 10.49 s (claimed 10–12 s). Three scans take 31.5 s (claimed ≥30 s). The <4 GB VRAM claim is plausible with uint8 storage, at ~2.7 GiB estimated (check 13).
Detailed processing description
Methods 6.3–6.11; Data availability
Methods 6.11 gives the full particle funnel for every dataset. The funnels are monotone and agree with the Results for four of five datasets (check 14). It also gives voxel-size corrections against reference structures (6SAE, 8RQB, 6HTX, EMD-4905), all of which match their cited sources (check 17), and hyperparameter choices. All five maps have EMDB accessions, although they are still on hold (check 16).
Q
H
C1>0
C1=−500
C1=+500
Q
∥U−Q∥=2.83
∣A1∣>∣C1∣
∣H−1∣
U
det=−1
R=HG
Q=I
Workflow paper without code, raw data or accessible maps
major · Data availability (L1422–1428); Methods 6.8; Suppl. Table 1 (not provided)
The main contribution is a 'complete workflow', yet TorchSlice and the preprocessing code are not released, and there is no code-availability statement. No raw 4D-STEM data for the new datasets are deposited (the only EMPIAR ID is the reused EMPIAR-12236). All EMD/PDB entries are still on hold (WAIT/HPUB/AUTH, check 16). Key per-dataset parameters (per-scan dose, defocus, CTEM dataset) sit in the Supplement, which was not part of the submission. As a result, the scan-geometry estimates (pitch 19.6/19.8 Å, 90.15°, −0.42°; check 19), the 0.71 Å model RMSD (check 18) and the iDPC/CTEM comparisons (check 20) cannot be reproduced. Fix: release code with a test dataset, deposit at least the dose-fractionated TMV 4D-STEM data in EMPIAR, and map each accession to its structure.
The dose-fractionation gain is not measured in a controlled experiment and the Results cite the wrong comparator
major · Sec. 2.7 L544–551; Methods 6.11.1 L1319–1326, 6.11.5
L548–551 attributes 'approximately 0.4 Å' to dose fractionation by comparing a 3.3 Å map from 5,330 particles with a 3.7 Å map from 3,607 particles. The two maps come from different grids, doses and defoci. Our ResLog estimate (check 03, B = 220/194 Ų) shows that the extra particles alone explain ~0.09 Å of this gap. The claim does survive against the full 5,683-particle single-scan map reported only in Methods 6.11.1 (≈0.41 Å at equal N). However, the Results should cite that comparator. The 5,622 → 5,330 particle removal is also undocumented (check 14). The obvious control is missing: sum the three fractions of the same micrographs into one non-fractionated reconstruction (or reconstruct from the summed 4D data), then refine the same particles. Without it, the size of the fractionation benefit, a headline claim, rests on cross-dataset comparison.
HBV resolution reported as both 4.7 Å and 4.5 Å
major · Sec. 2.6.1 L484; Fig. 4 caption; Fig. 4e legend; Methods 6.11.3 L1372
The Results (L484) and the Fig. 4 caption ('3.7, 4.1, 4.7 and 3.7 Å') give 4.7 Å. The Fig. 4e legend and Methods 6.11.3 (L1372) give 4.5 Å. The plotted curve first crosses 0.143 at 4.51 Å (check 01). It is noisy and recrosses between 4.51 and 4.37 Å. This is a confirmed inconsistency in a main result. The data support 4.5 Å, but readers cannot tell which map was deposited. Fix: state one value consistently, and consider a smoother FSC (more particles or masking) given the repeated crossings.
Automated model building is claimed without standard model validation
major · Sec. 2.7 L577–596; Fig. 4h
The abstract highlights 'automated building of atomic models', but the only evidence is a ChimeraX matchmaker RMSD of 0.71 Å to 6SAE. The paper gives neither the number of pruned versus total Cα pairs nor an all-pairs RMSD. It reports no MolProbity, Q-score, EMRinger or map–model FSC, and no ModelAngelo sequence-assignment statistics. Matchmaker's pruned RMSD can be very optimistic. Because the model is unreleased, we could not recompute it (check 18). Fix: report the full validation table that is standard for cryo-EM depositions.
Per-particle height analysis overstated
minor · Sec. 2.5 L386–406; Fig. 5c–d
L388 says the reference-based MSE has 'a single, unique minimum'. The extracted Fig. 5c curve has 6 local minima on the 4 nm grid, and the value at −8 nm is within 2×10⁻⁵ of the global minimum at −16 nm (check 09). The basin is broad, with roughly ±8 nm precision. L402–406 calls rod heights of +20 nm and −12 nm 'consistent with the known TMV diameter of 18 nm', but the 32 nm separation is 1.78 diameters. Rods stacked directly on one another would be 18 nm apart. Fix: report per-particle defocus uncertainties, for example from the curvature of the MSE or from repeat estimates, and either explain the 14 nm excess (a gap, a calibration factor, or the defocus-to-height scaling) or soften the claim.
iDPC and CTEM comparisons are confounded or unspecified
minor · Sec. 2.6.1 L465–477, L491–510; Fig. 4f
The iDPC comparison (9.0 Å vs 4.7/4.5 Å, L491–510) uses a different semi-convergence angle (2 vs 4 mrad) and a 4-segment detector. It is a single anecdote, yet it supports a general dose-efficiency conclusion. The CTEM dataset in Fig. 4f (instrument, dose, defocus, particle number) is not described anywhere in the main text (check 20). Fix: compare at matched SCA and dose, or restrict the conclusion, and describe the CTEM data.
No discussion of noise fitting in iterative ptychography at ~8 e⁻/Ų
minor · Sec. 2.3; Methods 6.8, L1196–1214
The reconstruction minimises an unregularised amplitude MSE for a fixed 100 iterations. The only justification for stopping is that the loss 'did not improve significantly' (L1201–1202), and learning rates are tuned to minimise the loss after 25 iterations. At these doses that criterion favours fitting noise. The paper does not show how the per-particle or map FSC depends on iteration number, nor how the noise model affects it. This matters for the fractionated data, where each scan has about a third of the dose. Fix: add an iteration-number or regularisation ablation on the final FSC.
'Fit resolution' τ is not a specimen-plane resolution
minor · Sec. 2.2 L209–211, L292–310; Figs 2d, 3d
τ (Å) is the Fourier conjugate of the Ronchigram angle, and larger values are better. It is reported as a 'resolution' (L209–211, Fig. 3d, τ ≈ 33 Å). At the Fig. 2 defocus, τ = 24/30 Å correspond to specimen periods of ≈6.9/5.2 Å, i.e. ring indices 3.5/5.8 (check 08). Readers used to CTF-fit resolution in Å will read it the wrong way round. Fix: rename τ (e.g. 'ring-detection limit') or also give the equivalent specimen period or ring number.
Inconsistent numbers for the headline TMV experiment
The total dose is given as 25 e⁻/Ų (L74–75), 3 × 8 = 24 e⁻/Ų (L95, Fig. 1a) and 23 e⁻/Ų (L532). The fraction labels add up to 8 + 7 + 8 (check 04). The per-scan dose for the single-scan datasets is never given. 'Helical pitch of 1.42 Å … 1.40 Å' (L1327–1328) refers to the rise, which is 3 × 23/49 = 1.408 Å (check 10). The Fig. 6f layer-line peaks sit at 23.49/11.75 Å against labels of 23.1/11.5 Å, a factor of ≈1.014. This suggests the axis uses the uncorrected pixel size (check 10). The Fig. 6e '3.4 Å' marker is drawn at 3.53 Å (check 02). 'Not contributing beyond 4.0/6.7 Å' (L629–635) corresponds to residual weights of 7.4%/6.1%, not zero (check 11).
Citation errors, including a reversed pairing from [53]
minor · Sec. 2.7 L559–565; References [3], [15], [39], [41], [43], [53], [54], [61], [65]
L561–565 pairs the CTEM motion of 4.2/2.9 Å with exposures '(1 s and 4.4 s respectively)'. In [53] the 4.2 Å motion belongs to the 4.4 s K2 exposure and the 2.9 Å motion to the 1 s Falcon II exposure (check 15). The paper's argument is actually better supported by the correct pairing. Further issues: [3]/[61] and [15]/[65] are duplicates. [41] cites a chapter in a TensorFlow textbook for PyTorch instead of Paszke et al. (2019), with the publisher printed as '???'. [39] is listed as 'in print' but is published. Author names in [43] and [54] are garbled.
Methods 6.11.2 (L1352–1356) says both 'the highest resolved class' and 'the largest class containing 4999 particles' was selected. Methods 6.6 says the Zernike-fit weights 'equal the variance' of the centre of mass, where inverse variance is presumably meant. Methods 6.8 writes 'fully coherent object' where 'probe' is presumably meant. The Fig. 6 caption refers to a white arrow in panel (f), which shows a black arrow, and omits e⁻ from the dose units. There are also typos ('transmisssion', 'opitimized', 'miminzing', 'botttom', 'polymonials'). The prior-work range '5.8 and 8.4 Å' (L43) omits the 12.1 Å Phi92 capsid map in [25].
−1
∣
∣A1∣>∣C1∣
What resolution do you get from the same 5,330 TMV particles when the three fractions are summed (or the 4D data are reconstructed as a single 23 e⁻/Ų scan) and processed identically? This would isolate the fractionation gain.
Why were 292 particles removed between the 5,622 in Methods 6.11.5 and the 5,330 used for the 3.3 Å map, and how does the map change if they are kept?
For the 0.71 Å RMSD to 6SAE, how many Cα pairs were retained after pruning, what is the all-pairs RMSD, and what are the MolProbity / Q-score / map–model FSC values of the ModelAngelo model?
What is the precision of per-particle defocus (e.g. repeat estimates or MSE curvature), and how do you reconcile the 32 nm rod separation in Fig. 5d with the 18 nm TMV diameter?
Which pixel size was used for the Fig. 6f frequency axis? The peaks sit at 23.49/11.75 Å rather than the labelled 23.1/11.5 Å.
How do the final FSCs change with the number of ptychographic iterations, or with a regulariser, at ~8 e⁻/Ų per scan?
Which instrument, dose, defocus and particle number were used for the CTEM dataset in Fig. 4f?
Will TorchSlice, the preprocessing code and the raw 4D-STEM data be released, and which EMD/PDB accession corresponds to which structure?
TMV voxel size corrected by matching the 'found helical pitch of 1.42 Å to the known helical pitch of 1.40 Å'; layer lines in Fig. 6f at 23.1, 11.5, 7.7 and 5.8 Å
(a) Terminology: 1.40 Å is the helical rise per subunit (3 × 23/49 = 1.408 Å), not the helical pitch (about 23 Å). Using 1.40 instead of 1.408 changes the voxel correction by 0.58% (3.3 Å → 3.32 Å), which is negligible. (b) The claimed layer lines are harmonics of 23.1 Å (23.10/11.55/7.70/5.78), which is consistent. (c) In the plotted Fig. 6f line scans, the two strong peaks sit at 23.49 Å and 11.75 Å in all four curves, and the arrows are drawn at 23.52, 11.74, 7.81 and 5.87 Å. The labels say 23.1, 11.5, 7.7 and 5.8 Å, a 1.7–2.2% offset. This is close to the 1.42/1.40 = 1.014 voxel correction (23.1 × 1.014 = 23.43 Å), suggesting the Fig. 6f axis uses the uncorrected pixel size while the labels use corrected values. The 7.7 and 5.8 Å lines are not resolved as peaks above the noise in the extracted curves (the paper calls them 'faint').
Computed the TMV helical parameters (pitch 23.0 Å, 49 subunits per 3 turns), the correction factors and the harmonics. Extracted the four Fig. 6f line scans and the arrow markers from the PDF vector paths (tick-calibrated, 0.025 Å⁻¹ ticks) and located the peaks.
Empirical dose weights (Fig. 6g): low frequencies retained from all fractions; the second fraction does not contribute beyond 4.0 Å, the third provides no information beyond 6.7 Å
Low-frequency weights at DC are 0.363 / 0.336 / 0.300 for the three fractions, so 'retained from all fractions' is confirmed. The cut-offs are approximate: fraction 2 still has weight 0.074 at 4.0 Å, falls below 5% only at 3.60 Å and below 1% at 2.89 Å. Fraction 3 has weight 0.061 at 6.7 Å, falls below 5% at 6.28 Å and below 1% at 4.78 Å. The stated limits therefore correspond to a weight of about 6–7%, not zero. The qualitative ordering is correct, but 'not contributing beyond' overstates the cut-off sharpness, and the paper gives no threshold. This is a minor issue.
Extracted the three Fig. 6g weight curves from the PDF vector paths (tick-calibrated, x linear in 1/d), evaluated the weights at 4.0 and 6.7 Å, and found where each drops below 5%, 2% and 1%.
Depth of focus of 221 nm at a semi-convergence angle of 4 mrad (300 kV)
λ = 1.9687 pm. 1.8λ/α² = 221.5 nm (prefactor needed for 221 nm: 1.796); the common 1.77λ/α² convention gives 217.8 nm (1.4% off). λ/α² = 123 nm and 2λ/α² = 246 nm do not match. The value corresponds to the ≈1.8λ/α² definition. The convention is not stated, but the order of magnitude supports the argument that offsets of a few tens of nm (Fig. 5d) are well inside the depth of focus.
Relativistic λ at 300 kV and standard depth-of-field conventions c·λ/α².
A 1024×1024 scan at 100 kHz takes 10–12 s; STEM movies take 'at least 30 s'; fewer than 4 GB VRAM suffice for 512×512 scan positions with uint8 CBEDs
1024²/100 kHz = 10.49 s (claimed 10–12 s). Three scans = 31.5 s (claimed ≥30 s). uint8 data for 512² positions = 2.25 GiB. Object plus optimiser state (3 × 0.125 GiB) plus minibatch buffers gives about 2.7 GiB in total, below 4 GB. The same data in float32 would need about 9 GiB, so the uint8 storage is what makes the <4 GB claim hold. This is an order-of-magnitude estimate only; actual TorchSlice memory use could not be measured because the code is not released.
Arithmetic on frame time and the memory footprint of the dominant arrays (96×96 binned Arina CBEDs, an object of about 4k² complex64 per 512² sub-scan, minibatch 32 with a 192² window).
Particle numbers are consistent between Results and Methods: TMV 3607, horse ApoF 4999, HBV 4052 → 4049, EMPIAR 22,420 → 22,198 (published 11,552), DF-TMV 5622 → 5330; voxel corrections 0.747 → 0.730 Å and 1.42 → 1.40 Å
Funnels are monotone, and the final counts agree between Results and Methods for TMV (3607), horse ApoF (4999), HBV (4049) and EMPIAR (22,198). Inconsistencies: (1) DF-TMV: Methods 6.11.5 ends at 5,622 particles. The 5330 used for the 3.3 Å map occurs only in the Results (L548), so the removal of 292 particles (5.2%) is undocumented. (2) Horse ApoF Methods L1352–1356 says both 'the highest resolved class was used for further refinement' and 'The largest class containing 4999 particles was selected', which is ambiguous about the selection criterion. (3) TMV Methods L1319–1326: the full 5,683-particle set already gave 3.7 Å, and the 3,607 'high per-particle scale' subset also gave 3.7 Å; the Results report only the subset. Voxel corrections rescale resolutions by 0.977 (EMPIAR) and 0.986 (TMV). The published EMPIAR 5.8 Å would become 5.67 Å at the corrected voxel size, so part of the stated 5.8 → 3.7 Å gain is calibration.
Rebuilt manuscript lines from the margin numbers, extracted the funnels, checked monotonicity and text/Methods agreement, and computed the voxel-correction scale factors.
Baselines from cited works: ptychographic SPA '5.8 and 8.4 Å' and TMV '6.4 Å' [25]; EMPIAR-12236 initially 5.8 Å from 11,552 particles [25]; iDPC TMV 'similar resolution' [18]; dose-fractionated CTEM TMV 3.4 Å [53]; CTEM TMV below 2 Å [54]; CTEM motion 4.2 and 2.9 Å at about 20 e−/Ų with acquisition times '(1 s and 4.4 s respectively)' [53]
Confirmed: [25] reports apoferritin at 5.8 Å from 11,552 particles, the Phi92 sheath at 8.4 Å (capsid 12.1 Å, omitted from the stated range), and a RELION ptychography reconstruction from 2120 particles at 6.4 Å in the TMV section. [18] reports iDPC TMV at 3.5 Å. [53] reports 3.35 Å (rounded to 3.4) and mean translational changes of 4.2 Å (K2 Summit) and 2.9 Å (Falcon II) at 20 e−/Ų. [54] reports 2.0 and 1.9 Å. EMPIAR-12236 exists ('Diffraction patterns of Apoferritin particles'). Discrepancy: in [53] the Falcon II exposure was 'one second' and the K2 Summit exposure 4.4 s, so the 4.2 Å motion belongs to the 4.4 s exposure and the 2.9 Å motion to the 1 s exposure. The paper's '4.2 Å and 2.9 Å … (1 s and 4.4 s respectively)' (L561–565) reverses this pairing. The argument that longer acquisitions give more motion actually holds in [53] (longer K2 exposure, larger shift), but the sentence states it incorrectly.
Fetched the abstracts and open full texts from the Europe PMC REST API (PMC11401879, PMC9467914, PMC4416312, PMC6831999) and the EMPIAR-12236 entry, and printed every sentence with resolutions, particle numbers, motion or exposure times.
Maps and models deposited as EMD-100046/PDB-34NO, EMD-100057/PDB-34OI, EMD-100042, EMD-100078/PDB-34OZ, EMD-100063
All five EMD accessions exist but are unreleased: EMD-100046, -100057 and -100078 are WAIT, -100042 HPUB, -100063 AUTH, with titles suppressed and no release date. PDB 34NO, 34OI and 34OZ are UNRELEASED (WAIT). The deposition claim holds, but none of the maps or models can be inspected yet. The text gives no mapping from accession to structure, no code availability statement, no repository link for TorchSlice, and no EMPIAR deposition of the new raw 4D-STEM datasets (the only EMPIAR ID is the reused EMPIAR-12236).
Queried the EMDB REST API for each EMD accession and the RCSB holdings API for each PDB ID, and searched the paper text for code-availability, repository and EMPIAR statements.
Reference structures used for model comparison and voxel calibration come from the cited works: PDB 6SAE (TMV FASTA and RMSD, [54]), 8RQB (mouse ApoF voxel size, [25]), 6HTX (HBV voxel size, [83]), EMD-4905 (horse ApoF initial map)
6SAE: 'Cryo-EM structure of TMV in water', 1.9 Å, primary citation Weis et al. EMBO Rep 2019 (= [54]). 8RQB: mouse heavy-chain apoferritin, 1.09 Å, primary citation Küçükoğlu et al. Nat Commun 2024 (= [25]). 6HTX: WT HBV core protein capsid, 2.66 Å, J Mol Biol 2018 (= [83]). EMD-4905: horse spleen apoferritin. All match the paper's usage. Note: the reported 0.71 Å RMSD to 6SAE could not be checked because the deposited model (PDB 34NO/34OI/34OZ) is unreleased.
Queried the RCSB data API (entry title, method, resolution, primary citation) and the EMDB API.
ModelAngelo model of the dose-fractionated TMV map fits PDB 6SAE with an RMSD of 0.71 Å (ChimeraX matchmaker)
The model (PDB 34NO/34OI/34OZ, all UNRELEASED per check 16) and the map are not accessible. The paper gives neither the number of pruned Cα pairs nor the all-pairs RMSD, so the 0.71 Å figure cannot be put in context.
Would require the deposited map and model (EMD/PDB on hold) to rerun matchmaker and report pruned vs all Cα pairs, plus standard validation (MolProbity, Q-score, map–model FSC).
Scan geometry from averaged Ronchigram cross-correlations: pitch 19.6 / 19.8 Å, scan axes at 90.15°, rotation −0.42° (Fig. 3b), and the power-spectrum fit resolution τ ≈ 33 Å from the cross-correlation peaks (Fig. 3c–d)
The principle behind these estimates is verified by simulation (checks 06–07), but the specific values cannot be reproduced without the raw data and code. The sign/orientation ambiguity of the polar decomposition (check 07) cannot be checked against the reported −0.42° rotation either.
Would require the raw 4D-STEM dataset (not deposited) and the TorchSlice code (not released).
iDPC on the same HBV grid gives 9.0 Å with the same number of particles (at 2 mrad SCA instead of 4 mrad); Fig. 4f frequency-resolved correlation of ptychography vs CTEM
No acquisition parameters for the CTEM dataset in Fig. 4f and no iDPC FSC curve appear in the main text. The supplement (Suppl. Fig. 6, Suppl. Table 1) was not part of the uploaded material.
Would require the iDPC and CTEM datasets, or at least their acquisition parameters (dose, defocus, particle numbers), which are in the unavailable supplement or not given at all.
FSC=0.143 resolutions of standard ptychographic SPA: TMV 3.7 Å, horse ApoF 4.1 Å, HBV 4.7 Å (L484, Fig. 4 caption) / 4.5 Å (Fig. 4e legend, Methods L1372), mouse ApoF 3.7 Å
TMV: claimed 3.7, obtained 3.69 Å. Horse ApoF: claimed 4.1, obtained 4.14 Å. Mouse ApoF: claimed 3.7, obtained 3.63 Å (all within ±0.1 Å). HBV: the plotted curve crosses at 4.51 Å, matching the legend and Methods (4.5 Å), but the Results text (L484, 'with 4.7 Å resolution was obtained from 4049 particles') and the Fig. 4 caption ('3.7, 4.1, 4.7 and 3.7 Å') state 4.7 Å, a 0.19 Å mismatch. The HBV curve is noisy, with repeated crossings between 4.51 and 4.37 Å.
Extracted the four vector FSC curves from the PDF (p. 8), calibrated axes on tick marks (residual <1e-4), and found the first downward 0.143 crossing. The drawn threshold line sits at FSC 0.1429.
Dose-fractionated TMV: 3.5 Å after patch motion and 3.3 Å after reference-based motion correction (Fig. 4i); single fractions 3.4 / 3.9 / 4.4 Å and combined 3.3 Å (Fig. 6e)
Fig. 4i: patch motion claimed 3.5, obtained 3.53 Å; reference-based claimed 3.3, obtained 3.31 Å. Fig. 6e: dose-fractionated claimed 3.3, obtained 3.34 Å; 0–8 e−/Ų claimed 3.4, obtained 3.44 Å; 8–15 claimed 3.9, obtained 3.85 Å; 15–23 claimed 4.4, obtained 4.40 Å. All within ±0.1 Å. Minor annotation slip: the Fig. 6e vertical marker labelled '3.4 Å' is drawn at 3.53 Å (the curve crosses at 3.44 Å). The other markers sit at 3.34, 3.86 and 4.39 Å.
Vector FSC curves extracted from PDF pp. 8 and 11. Axes were calibrated on tick marks, curves were matched to legend entries by legend-line colour, and the first 0.143 crossing was found. Vertical resolution markers were also converted to Å.
Dose fractionation improves TMV resolution 'by approximately 0.4 Å' (3.7 Å single scans → 3.3 Å with 5330 particles). Mouse ApoF beats horse ApoF because of particle number. The EMPIAR-12236 gain 5.8 → 3.7 Å (11,552 → 22,198 particles) is partly due to the new preprocessing.
(a) Against the 3,607-particle map cited in the Results, the extra particles (→ 5330) explain 0.09 Å (22%) of the gain, leaving about 0.33 Å at equal N. But Methods 6.11.1 shows the full 5,683-particle single-scan set also reached 3.7 Å. Scaled to 5,330 particles, that gives 3.715 Å, so the dose-fractionated gain at equal particle number is about 0.41 Å, consistent with 'approximately 0.4 Å'. The Results should cite this comparator. Different grids, doses and defoci remain uncontrolled and cannot be tested from the paper. (b) Horse vs mouse ApoF: the implied B is 219.9 Ų, matching the reported ResLog B of 220 Ų, so the particle-number explanation is quantitatively consistent. (c) EMPIAR-12236: particle doubling alone explains only 5.8 → 5.1–5.35 Å (21–33% of the gain for B = 150–250 Ų; the implied B if particles alone were responsible would be 30 Ų), so most of the gain comes from processing, as claimed. The 0.747 → 0.730 Å voxel correction alone accounts for 5.8 → 5.67 Å.
Rosenthal–Henderson/ResLog relation 1/d² = (2/B) ln N + c, using the paper's own B-factors (220 Ų single scans, 194 Ų dose-fractionated). Predicted resolution changes from particle number alone and the implied B-factors, for both single-scan comparators reported in Methods 6.11.1 (3,607 and 5,683 particles, both 3.7 Å).
Total dose of the dose-fractionated acquisition: 25 e−/Ų (L74–75), 3 × 8 e−/Ų (L95, Fig. 1a), 23 e−/Ų (L532); fractions 0–8, 8–15, 15–23 e−/Ų (Figs 4g, 6)
Three different totals for the same acquisition: 25 (L74–75), 24 (= 3 × 8, L95) and 23 e−/Ų (L532; fraction labels 8 + 7 + 8 = 23). The second fraction is 7 e−/Ų, not 8. The Fig. 6 caption gives the fractions as '0-8, 8-15, 15-23 Å2' (missing e−). The per-scan dose for the single-scan datasets (TMV, horse ApoF, HBV) appears nowhere in the main text (Suppl. Table 1 is not provided). For context, 8 e−/Ų at a 5 Å step and 100 kHz implies 200 e− per position, i.e. about 3.2 pA.
Regex extraction of all dose statements from the paper text, plus recomputed sums and per-fraction doses.
Eq. (2): the power spectrum of out-of-focus Ronchigram shadow images follows rCTF(τ) = E(τ) sin(π/λ τᵀH⁻¹τ), with H from Eq. (1); zeros at τ_n = sqrt(nλC1), elliptical under astigmatism
Pure defocus: minima at 8.85, 12.54, 15.39, 17.78, 19.88 Å vs predicted 8.87, 12.55, 15.37, 17.75, 19.84 Å (max error 0.27%, tolerance 3%). Astigmatic case: the elliptical average with the true H gives minima at s = 0.98, 2.00, 3.02, 4.01, 5.00 (max deviation 0.019). Ignoring A1 washes out the rings (contrast 0.51 vs 0.24). Axial minima match sqrt(nλh_i) within 1.7%. Side result relevant to check 07: C1 = −4000 Å gives identical zeros, so the power-spectrum fit cannot determine the sign of the defocus.
Wave-optical numpy simulation: 300 kV, α = 10 mrad, C1 = 4000 Å (±), A1 = 1000 Å. Used 60 random weak-phase objects and far-field intensities minus the object-free disk, apodised in the BF disk, Fourier transformed over k, with power spectra averaged. Minima were located radially, in ±8° axial wedges, and in an elliptical average over s = τᵀH⁻¹τ/λ (Eq. 2 predicts zeros at integer s).
Neighbouring Ronchigrams are shifted by the shadow step Δθ = H⁻¹Rs (Eq. 8). The cross-correlation peak is surrounded by the IFT of the shadow-image power spectrum, so Fourier transforming the peak-centred cross-correlation gives a power spectrum to which the rCTF can be fitted.
Peak lags match ±H⁻¹s/λ within 0.17–0.37% in all six cases, including the astigmatic one (e.g. −0.05065 vs −0.05079 Å⁻¹). The power spectrum recovered from the cross-correlation has minima at 8.85, 12.57, 15.37, 17.77, 19.89 Å vs predicted sqrt(nλC1) = 8.87, 12.55, 15.37, 17.75, 19.84 Å (max error 0.29%), identical to the direct power spectrum. The lag direction reverses for C1 < 0, so the shift carries the defocus sign only if the scan rotation R is known independently (see check 07).
Wave-optical simulation (300 kV, α = 10 mrad; C1 = ±4000 Å and an astigmatic case). The object was shifted by a 5 Å scan step in two directions, and windowed difference Ronchigrams were cross-correlated over 40 objects. Sub-pixel peaks were compared with H⁻¹s/λ (one fixed global sign set by the simulation's FFT/shift convention). The peak-centred cross-correlation was then Fourier transformed and its radial minima located.
Eq. (11): G R⁻¹ = Q H⁻¹, from which Q and H⁻¹ 'can then be separated using polar decomposition' into a unitary orientation matrix and the symmetric inverse defocus-astigmatism matrix
Exact recovery only for positive-definite H⁻¹ (C1 = +500 nm: error 0.000). For negative defocus (the regime of most datasets, e.g. TMV −190…−800 nm, ApoF −220…−920 nm per Methods 6.11), polar decomposition returns C1 = +500 nm instead of −500 nm and U rotated by 180° from Q (||U − Q|| = 2.83). For |A1| > |C1| it returns |H⁻¹| (e.g. C1 = +150 instead of +100 nm) and an improper U (det = −1), which is not a rotation. The procedure as described therefore cannot fix the defocus sign; an extra constraint is needed (known camera orientation, or the sign from another source), and the paper does not state one. The rCTF power spectrum cannot provide the sign either (check 05). Separately, Eq. (10) R = HG contradicts Eq. (11) whenever Q ≠ I (||HG − R||/||R|| = 0.30 for a 0.3 rad rotation), because Eq. (8) omits the scan–camera rotation.
Built A = G R⁻¹ = Q H⁻¹ synthetically for 200 random rotations Q per case (positive defocus, negative defocus at −500 and −190 nm, and |A1| > |C1|), applied scipy.linalg.polar (A = U P), and compared U, P⁻¹ with Q, H. Also tested Eq. (10) against Eq. (11).
Local defocus estimation raises the CC = 0.25 ring-detection limit 'by 25 % from τ = 24 Å (global) to τ = 30 Å (local)'; the local defocus landscape gives a foil tilt of 17.4° (stage tilt 20°)
CC drops below 0.25 at τ = 24.4 Å (global; claimed 24) and 29.7 Å (local; claimed 30); (30−24)/24 = 25.0% (from the extracted values, 22%). rCTF zeros give λC1 = 165.1 Ų (C1 = 0.839 µm, global) and 156.0 Ų (0.793 µm, local), with intercepts of 5.2 and 1.0 Ų, consistent with the Fig. 2e range 0.72–0.93 µm. Fig. 2e patch C1 spans 0.722–0.914 µm (ΔC1 = 0.193 µm), so a planar tilt lies between 14.9° and 20.6°, which contains the claimed 17.4°. Interpretive note: the τ 'fit resolution' is not a specimen-plane resolution. τ = 24/30 Å correspond to specimen periods of about 6.9/5.2 Å (ring indices 3.5/5.8) at this defocus, which the text does not explain.
Extracted the Fig. 2d CC step curves and red rCTF curves from the PDF vector paths (tick-calibrated) and found the CC < 0.25 point. Fitted the rCTF zeros to τ_n² = nλC1. Converted τ to specimen-plane period d = λC1/τ. Mapped all 1849 Fig. 2e surface-patch fill colours through the embedded colour-bar image to C1 and bounded the planar tilt over a 1024 × 5 Å = 0.512 µm field.
In Fig. 5c the MSE 'reveals only a single, unique minimum' about 20 nm below the initial defocus, while the loss has several local minima and is 'significantly lower at +80 nm'. Vertical rods at about +20 nm and diagonal rods at −12 nm are 'consistent with the known TMV diameter of 18 nm'.
Confirmed parts: the curves are sampled every 4.00 nm (Methods: ±100 nm, 51 steps; 50 plotted points). The MSE global minimum is at −16 nm ('approximately 20 nm lower'). The loss at +80 nm (1.46080) is below the loss at the MSE minimum (1.46248), but by only 0.115% of the loss value, and the loss keeps decreasing to the edge of the scanned range (global minimum at +100 nm), i.e. it has no interior optimum. Discrepant parts: (a) the MSE curve has 6 local minima on the grid (−28, −16, −8, +16, +32, +40 nm), and −8 nm is within 2×10⁻⁵ of the global minimum. The basin is broad and unimodal overall, but 'single, unique minimum' overstates the precision (an uncertainty of about ±8 nm is visible). (b) The rod heights +20 and −12 nm give a 32 nm separation, which is 1.78× the 18 nm TMV diameter. Rods resting on each other would be 18 nm apart centre to centre, so the stated consistency needs a 14 nm gap or a calibration factor that the paper does not discuss.
Extracted the Fig. 5c loss and MSE curves from the PDF vector paths (tick-calibrated: x every 25 nm, both y axes from tick marks), located local and global minima, compared loss values, and computed the implied height separation.
Exists and supports the claims (TMV at 2.0 and 1.9 Å; source of PDB 6SAE). Formatting: the PDF prints the third author as 'Der Hocht, I.' instead of 'von der Hocht, I.'
A1
∣
>
∣C1∣
Add a same-data control for dose fractionation (summed fractions, same particles, same processing), and cite the 5,683-particle single-scan map in the Results.
Release TorchSlice and the preprocessing code with a small test dataset. Deposit the dose-fractionated TMV raw 4D-STEM data in EMPIAR.
Report a full model-validation table (MolProbity, Q-score, map–model FSC, pruned/total Cα pairs) for the ModelAngelo model.
Make the HBV resolution (4.5 vs 4.7 Å) and the total dose (23/24/25 e⁻/Ų) consistent throughout, and give per-scan doses for all datasets in the main text.
Rename or explain the τ 'fit resolution' and give the equivalent specimen period or ring number.
Give per-particle defocus uncertainties and revisit the height argument in Fig. 5d.
Check the Fig. 6f frequency axis against the corrected pixel size, and state the weight threshold behind the 4.0/6.7 Å cut-offs in Fig. 6g.
Compare iDPC and ptychography at a matched semi-convergence angle and dose, and describe the CTEM dataset in Fig. 4f.
Fix the reference list: duplicates [3]/[61] and [15]/[65], the PyTorch citation [41], the reversed exposure pairing from [53], and the [39]/[43]/[54] metadata.